LCR Meter 11021 / 11021-L User’s Manual
4-12
2. After all settings are done, press [MAIN INDEX] and [F2] (COUNT) to enter the
BINNING test screen as shown below:
The BIN 0 counter is to calculate the number of tested secondary parameters that exceed
the HI/LOW setting.
The BIN 1 counter in this example is to calculate the number of tested secondary
parameters that do not exceed the HI/LOW setting and the number of tested primary
parameters that are within the range of ±0.1%. The rest of BIN 2~8 is set accordingly.
The X counter is to calculate the number of tested secondary parameters that doe not
exceed the HI/LOW setting but the tested primary parameters exceed the range setting of
all BIN 1 ~ 8.
The T counter is to calculate the number of all tested items.
RESET: Press [F1] will clear all of the counters.
The H/L setting range for BIN1~BIN8 is –9999.99%~+9999.99%.
When exiting the BIN SETTING screen, check if the HI and LO setting parameters are
reasonable. Once HI < LO, the HI and LO settings will be switched automatically.
4.4.5 Setting COMPARE
1. When the meter is powered on normally, press [MAIN INDEX] to enter the main screen
as shown below:
2. Press [F2] to enter the screen as shown below:
< BIN SETTING > Zs-θ
INCREASE
4.H: +0000.40 % L: -0.40 % DECREASE
5.H: +0.50 % L: -0.50 % DIGIT
6.H: +0.60 % L: -0.60 %
F1
F2
F3
F4
< BIN SETTING > Zs-θ
INCREASE
7.H: +0000.70 % L: -0.70 % DECREASE
8.H: +0.80 % L: -0.80 % DIGIT
F1
F2
F3
F4
0: 0 4: 0 RESET
1: 32857 5: 0 8: 0
2: 1128 6: 0 X: 0
3: 253 7: 0 T: 34238
F1
F2
F3
F4
< MAIN INDEX > BINNING ÅSet the sorting test
COMPARE ÅSet HI/LO limit for comparison
CORRECTION ÅZeroing calibration
F1
F2
F3
F4