Description of Panel
4-11
Assuming to set PRI_NOMINAL to 10KΩ, SEC_HI_LIMIT to+0.5°, and SEC_ LO_ LIMIT
to -0.5°, follow the steps below for setting:
(1) Press [], [] to move the cursor to PRI_NOMINAL, and press [F3] or [], []to
move the cursor to the denary number, then press [F1] to adjust the number to 1.
Press [F3] or [], [] again to move the cursor to (-), and press [F1] to set the unit
to K.
(2) Press [] to move the cursor to SEC_HI_LIMIT, the cursor will stop at +. Then
press [F3] or [], [] to move the cursor to the first digit of decimal point, press [F1]
again (the number increased) to adjust the number to 5.
(3) When the previous setting is done, press [], the cursor will move the next setting
item SEC_LO_LIMIT automatically. Press [F1] to set it to (-), then press [F3] or [],
[] to move the cursor to the first digit of decimal point, press [F1] again (the
number increased) to adjust the number to 5.
(4) Now, the setting is done as the figure shown below.
< BIN SETTING > Zs-θ
INCREASE
PRI_NOMINAL: 10.000 KΩ
DECREASE
SEC_HI_LIMIT : 0.500 -°
DIGIT
SEC_LO_LIMIT : -0000.500 -°
F1
F2
F3
F4
Description: PRI_NOMINAL indicates the primary parameter. The setting range is
0000.0000p~9999.9999G.
SEC_HI_LIMIT indicates the comparison high limit value for secondary
parameter. The setting range is –999.9999G~+999.9999G.
SEC_LO_LIMIT indicates the comparison low limit value for secondary
parameter. The setting range is –999.9999G~+999.9999G.
When the HI/LOW limit setting for primary and secondary parameter is done, press
[] will appear the BIN SETTING screen (as shown below). The operation is same as
(1) ~ (3). If the HI/LOW setting of BIN is symmetry such as ±0.1%, then it is only
necessary to set the H value to +0.1% and press [TRIGGER], the L value will be set
to -0.1% automatically. If the HI/LOW setting of BIN is not symmetry, then set the H
value first and L value next.
< BIN SETTING > Zs-θ
INCREASE Å Increase the number
PRI_NOMINAL: 0000.0000 -Ω
DECREASE Å Decrease the number
SEC_HI_LIMIT: 0.0000 -°
DIGIT Å Move the cursor rightward
SEC_LO_LIMIT: 0.0000 -°
F1
F2
F3
F4
< BIN SETTING > Zs-θ
INCREASE
1.H: +0000.10 % L: -0.10 % DECREASE
2.H: +0.20 % L: -0.20 % DIGIT
3.H: +0.30 % L: -0.30 %
F1
F2
F3
F4