Description of Panel
4-7
3. Press [F1] to show the following screen:
4. Select Single or Multi frequency as need, and the screen appears as below:
The figure above tells the user that it will do the open correction now. Please prepare the
test cable and press [TRIGGER] to start the correction. Be sure to use the
Test Cable of
4BNC/2Clips shipped along with the LCR Meter and keep it in open state as the figure shown
below.
When there is an open zeroing test error, the display will show [FAIL] to indicate the test has
encountered a problem. Please check the test cable to see if it is disconnected or there is
any bad connection of the clip. Do the open zeroing test again after the examination.
[PASS] will appear on the display to indicate the open zeroing test is done when the test is
correct. Press [EXIT] in the following screen to quit.
< CORRECTION OPEN > SINGLE Å Single frequency
MULTI Å Multi frequency
ABORT
F1
F2
F3
F4
Å Cancel correction, return to
previous page
< CORRECTION OPEN >
OPEN CIRCUIT TEST LEADS
THEN PRESS TRIGGER.
ABORT
F1
F2
F3
F4
Å Cancel correction, return to
previous page
< CORRECTION OPEN >
F: 10KHz Cp: 0.00 pF
PASS
Press ‘EXIT’ to exit. EXIT
F1
F2
F3
F4
Å Quit the screen