LCR Meter 11021 / 11021-L User’s Manual
5-14
(8) Command: SYSTem:ALARm
Parameter: {PULSE | CONTINUOUS}
Return: {PULS | CONT}
Function: It sets the beep to PULSE or CONTINUOUS.
It is used when doing BIN SORTING and COMPARE.
Syntax: SYST:ALAR PULS
Query Syntax: SYST:ALAR?
Return Data: PULSE
(9) Command: SYSTem:HANDler
Parameter: {CLEAR| HOLD}
Return: {CLEA | HOLD}
Function: It sets to HANDLER MODE, see section 4.3.1.
Synt
ax: SYST:HAND CLEA
Query Syntax: SYST:HAND?
Return Data: CLEAR
TRIGger Command Set
(1) Command: TRIGger:DELay
Parameter: The trigger delay time.
Unit: S.
Return: The trigger delay time, the format is < NR3 >
Function: It sets or queries the trigger delay time in the range of 0~9.999 and only
valid in EXT TRIG MODE.
Syntax: TRIG:DEL 1.000 Set the time to delay 1.000S for trigger.
Query Syntax: TRIG:DEL?
Return Data: 1.000000e+00
(2) Command: TRIGger[:IMMediate]
Parameter: None
Return: None
Function: It triggers the measurement no matter what the present state is.
Syntax: TRIG
(3) Command: TRIGger:SOURce
Parameter: {BUS;MANual |EXTernal|INTernal }
Return: { MAN.|EXT.|INT. }
Function: It sets or queries the present trigger mode.
Description: BUS/ MANual Triggered by bus/manual
EXTernal Triggered externally
INTernal Triggered internally
Syntax: TRIG:SOUR BUS
Query Syntax: TRIG:SOUR?
Return Data: MAN.
(4) Command: TRIGger:EDGE
Parameter: {FALLing | RISIng}
Return: {FALLING | RISING}
Function: It sets the TRIGGER mode to EXTERNAL, RISING or FALLING edge
triggering.
Syntax: TRIG:EDGE FALL