LCR Meter 11021 / 11021-L User’s Manual
5-20
Execution Error
0 No error
There is no error message exist at present.
-102 Syntax error
Invalid character exists in the command string, ex. TST?
-104 Data Type error
The parameter is not defined in the command string.
-106 Illegal parameter value
The parameter type is error in the command string.
-202 Setting conflict
One of the following conditions may cause the error:
Sending out *TRG or TRIGGER command when the trigger mode is external.
-203 Data out of range
The data parameter exceeds the valid range, ex. TRIGger:DELay 10
-211 Data stale
The following condition may cause the error:
-224 Self-test failed
The self-test executed via remote interface (*TST) is failed. In addition, there are
other test errors. Refer to *TST? command for the description of return format.
-225 Too many errors
More than 16 errors are occurred and the error queue is full. It will not store other
errors until some of them are deleted. The error queue will be cleared after
powered off or executing *CLS (clear state) command.
-226 Query INTERRUPTED
When the device is in sending data state, the sending data is interrupted due to
device change to receiving state after got the new command. The output buffer
will be cleared.