Calibration
4-1
4. Calibration
4.1 Introduction
The ac source built a simple way to calibrate the output and measurement accuracy without
opening cover. Users can do it just follow the procedures step by step. A voltage meter,
current meter, suitable load and +5V dc source are needed for while calibration procedure.
Connections for these instruments please refer to Figure 4-1. There are three items need to
calibrate. But it is not necessary to calibrate all at once. User can just choose one item if
needs.
Figure 4-1
4.2 MANUAL CALI Functional List
Users can choose "4. MANUAL CALI " in CHOICE PAGE to enter the calibration procedure.
Before showing the calibration items, for safety reason, user must enter password. The
password is shown in this manual, in order to confirm the user read the manual before doing
the calibration procedure.