Programmable DC Electronic Load 6310 Series Operation & Programming Manual
1.3 Overview of Key Features
A. Mainframe
■ Flexible configuration using plug-in electronic load modules to mainframes.
■ Local operation from front panel keypad.
■ Remote control via GPIB or RS-232C interface.
■ Photo coupler isolation offers true floating Load.
■ Automatic fan speed control to reduce noise.
■ Up to 8 channels for one Mainframe.
B. Load
■ Constant current (CC), constant resistance (CR) and constant voltage (CV)
operation modes.
■ Programmable slew rate, load levels, load periods and conduct voltage (Von).
■ Programmable dynamic loading with speed up to 20KHz.
■ Minimum input resistance allows load to sink high current even with low input
voltage (1 V).
■ Selective voltage and current ranges.
■ Remote sensing capability.
■ 100 sets of memories to save/recall user-definable setups.
■ 10 sets of programs to link files for automatic test.
■ 15-bit A/D converter with precision measurement.
■ Short circuit simulation.
■ Master/Slave parallel control mode, allow synchronous load control under static and
dynamic loading mode
■ Automatic GO/NG inspection to examine if UUT within spec.
■ Independent GO/NG signals for each channel.
1.4 Specifications
Mainframe: 6334/6332
AC input: 115/230 interchangeable or 100/200 interchangeable Vac line
Fuse: 5A, 250V/2A, 250V
Amplitude: ±10%
Frequency: 47 to 63 Hz
Maximum VA: 180VA/100VA
Trigger output: Vlo = 0.8V maximum at Ilo = 1 mA
Vhi = 3.2V minimum at Ihi = −40μA
Weight: 24Kg/15Kg
Dimension:
Width: 440mm/275mm
Height: 177.4 mm (excluding foot stock)
186mm (including foot stock)
Depth: 560mm (including Load module)
The specifications of Load are listed from the next page on.
1-2