9.4. Adjusting scale for dosage procedures
During dosage substances essential matter is fast reaction of the balance on mass
increment. During this process digital filters must be turn off because they make
reaction time longer although they help if any disturbances occur.
Procedure
Press the TARE button, when horizontal lines appear on the display press the CAL
button
The balance shows first function in the menu. Press the F key repeatedly
until command FIL appears on the display.
Press the PRINT button
Pressing the F button sets up new value of the function FIL but comply
with following conditions:
FIL 0 – understood value of the digital filters
(recommended for weighing process)
FIL 1 – recommended for dosage, pouring, filling etc.
Confirm chosen value by pressing the PRINT button
Attention:
In the menu there is also function AUTO (autozero). This function controls zero
point of the scales. User can turn it off before the dosage. The procedure of turn
autozero off is included in the instruction manual.