15. ORDERS ON DIGITAL DISPLAY
Err- 5 - signalling terror of temperature conveter, to start scale once
again turn on and off power supply
FuLL – 1 - overcrossed range of conveter A/D
FuLL- 2 - overcrosed range of weighing
nuLL - too small load of scale (np. lack of scale)
tESt - autotest of scale is done (after turn power on)
CAL - calibration is done
STAB - definition parameter of stability for sedning data
DATE - printing data with date and time
CONT - constance transmission
AUTO - turn on/off autozero
- Co - - choose kind of work (checking thickness of solids)
- Li - - choose kind of work (checking thickness of liquids)
tE_L 00 - order to write temperature (during checking thickness)
H2O - destilled water (thickness of solids)
C2H5OH - alkohol - spirit (thickness of solids)
Spirit 100% +/- 0.1% in temp. 20
0
C
AnotHEr - another liquid with klown thickness(thickness of solids)
VOLUME - order to write in value of capacity of plunger (checking
thickness of liquids)
LoAD A -order to put sample on upper scale (checking thickness)
LoAD L - order to put sample on down scale (checking thickness)
Err-3 - too mach difference between start mass and settled factory
mass and real start mass of scale – there is possibility to
appears bigger errors
CAL Err - error in calibration (calibrate scale once again)
bod - function of setting speed of transmission
AC_t - automatic calibration of scale
AuE - average result
FIL - digital filters
St_uni - default unit of weighing