- 5 -
3. CONDITIONS OF APPROPIATE USING
scales should be used and stored in rooms without vibrations and shakes,
without drafts and dust
temperature of air should be +15
o
C ÷ +35
o
C
during using balances changes of temperature shouldn’t be higher than
0,5
o
C per one hour
scales should be replaced on wall console or stable table, far from sources
of heating
strong magnes is installed inside of balance – so it’s important to be carefull
during weighing magnetic materials. Use hang loads if it’s possible not to
expose magnetic load to magnes influence. Suspension is replaced on basic
of the balance.
If static electrisity has influence on scale annunciators it’s neccesary to
connect it to earth. Earth screw is replaced in back part of balance base.
Scale should be replaced in horizontal level to assure precision of weighing