6-1-2 LOW Setting Value (LO.SET)
Parameter name Setting value Remarks
LOW setting value (LO.SET)
Sets a LOW setting value for pass/fail judgment.
−199.9999 to 199.9999 (mm) [Default: 1.0000]
Operation procedure
Note
Configure settings such that HIGH setting value > LOW setting value.
Although it is possible to set HIGH setting value < LOW setting value, the GO judgment will not be
output.
Related information
The LOW setting value can always be displayed on digital display SUB (green).
For details, refer to "5-2 SUB Display Switching Mode."
HIGH and LOW setting values can be set by teaching using a master work, upper limit work, and
lower limit work.
For details, refer to "5-3 Teaching Mode."