Troubleshooting Guide
6
how hard the patient breathes; when the patient breathes weakly, the waveform’s amplitude seems
low.
The ECG module is defective.
Table 4-4
4.6.5 SpO
2
Failures
Symptoms
Possible Cause
Corrective Action
No waveform
and value
The SPO
2
probe is defective
Replace the SPO
2
probe
The SPO
2
probe and the SPO
2
module are mismatched.
Replace the SPO
2
probe that matches the SPO
2
module
The SPO
2
module is defective
Replace the SPO
2
module
Parameter board is defective
Replace the ECG module
There is large
deviation in
SPO
2
value and
waveform is
chaotic
The tested object’s fingers
move a lot.
Do not move fingers.
The SPO
2
probe is defective
Replace the SPO
2
probe
The SPO
2
module is defective
Replace the SPO
2
module
Possible causes of the symptoms:
Incorrect SPO
2
value and chaotic waveform
This symptom can also be caused by the object’s hypothermia, poor perfusion, shock, or the use of
vasoactive drugs, in addition to the previously mentioned causes.
Typically, a defective SPO
2
plate that produces no waveform data shows up as follows:
There is no lighting in the SPO2 probe light tube.
The probe tube lights up, but when a finger is inserted, the light flashes continuously.
Table 4-5
4.6.6 NIBP Failures
Symptoms
Possible Cause
Corrective Action
Can’t start air
inflation
The air valve or air pump is
defective
Check if the air valve or pump is defective. If the
machine makes a ticking noise when it is turned on,
the air valve is likely in good working order.
Parameter board is defective
Replace the Parameter board
The main board is defective
Replace the main board