EasyManua.ls Logo

Comen N Series - CO Failures; Software Upgrade Problems; Technical Alarm Messages

Default Icon
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting Guide
9
4.6.9 CO
2
Failures
Symptoms
Possible Cause
Corrective Action
Large CO
2
reading errors
CO
2
analyzer does not be reset
to zero
Zeroing must be done before use
The settings are incorrect
Select the standard ones based on local height: O
2
compensation, pambient, altitude, and equilibrium
gas, which can be set by referring to the
“Barometric Pressure Conversion Table” in the CO
2
part of the operating instructions.
CO
2
cannot be
measured
Operating status is not set
The operating status is set to “Measurement” and
the CO
2
switch is set to “On”.
The internal cables of the
instrument are poorly connected
or damaged
Re-plug or replace the cables
CO
2
cables are poorly
connected
Re-plug or replace the cables
The CO2 module is defective
Replace the CO
2
module
Table 4-9
4.6.10 Software Upgrade Problems
Symptoms
Possible Cause
Corrective Action
System upgrade
failure
The USB interface/ main board
is damaged
Replace the main board
The USB-stick is not plugged
in.
Check the USB-stick on the USB interface
Download the wrong upgrade
package
Use the correct upgrade package
Table 4-10
4.6.11 Technical Alarm Messages
Please refer to the User manual.

Table of Contents