Safety
2
explosion or fire from happening.
Please carefully install the power lines and the cables for various accessories to prevent the patient from
being constricted or suffocated or the cables from getting entangled and keep the patient free from
electrical interference.
Do not use a mobile phone near the monitor, because the mobile phone will generate a very strong
radiation field and disturb the functions of the monitor.
For the patient with a pacemaker, a cardio tachometer might measure the heart rate by the pulse of the
pacemaker during cardiac arrest or arrhythmia. Do not completely rely on the alarm of the cardio
tachometer. The patient with a pacemaker shall be closely monitored. For the inhibiting capacity of
relevant equipment on the pacemaker, refer to the Instruction Manual.
The operators shall not touch the patients, tables, and instruments during the defibrillation period.
Before reusing these cables, check whether the function is normal.
The equipment connected with the monitor shall form an equipotential body (the protective grounding
wire is effectively connected).
When the monitor is used in conjunction with the electro-surgery unit, the user (doctor or nurse) shall
ensure the patient's safety.
The physiological waveform, physiological parameters, and alarm information, etc. displayed by this
monitor shall be for the doctors’ reference only and cannot be directly used as the clinical treatment
basis.
The electromagnetic field will affect the performance of this instrument, so the use of the other
equipment near this instrument must meet corresponding EMC requirements. For example: Mobile
phones and X-ray or MRI equipment may be an interference source because they will transmit high-
strength electromagnetic radiation.
This is not a treatment device.
The ECG restore time following defibrillation shall not exceed 5 seconds, while the restore time for
other parameters shall not exceed 10 seconds.
Any conductive connections between the monitor’s sensors, rows of wires, and the HF surgical
equipment shall be avoided when the two are used simultaneously to protect the patient from burns.
Caution
Please properly install or move this instrument and prevent the instrument from being
damaged due to fall, collision, strong vibration or other external mechanical forces.
Before the instrument is switched on, please confirm whether the power supply used meets
the requirements for power supply voltage and frequency designated specified on the
nameplate label or in the instruction manual of this instrument.
When this instrument and its accessories are about to exceed the service life, they must be
disposed of according to local relevant laws and regulations or the rules and regulations of
the hospital.
Reusing disposable accessories more than once may cause performance degradation or
cross-contamination.
Please remove the batteries and store them appropriately if the monitor won’t be used for
an extended period.