52
Installation of the SPEEDY WATER demineralization system
To install the SPEEDY WATER demineralization system on a steriliser already fitted with
automatic water inlet 29 and direct drain 30, follow these steps:
• Install support bracket A to a wall suitable to hold the weight of the plate and of the
purification block B during operations (5.6 kg). Take into account the vertical space
needed to position the block.
• Turn off the mains water tap and remove the plugs from the connections behind the
bracket.
• Connect the pipe from the water mains to the rear right section of the bracket (bearing
the label “IN” at the front) and the pipe to the steriliser to the rear left section of the
bracket (bearing the label “OUT” on the front). Use the supplied pipes (2 m each) and
push them as far as possible into the connections.
• Remove the plugs positioned under the front cover of the bracket.
• Remove the protective plugs from the purification block B inlets and wet their seals with
demineralized water to ensure maximum efficiency.
• Lift the cover of the bracket and place the purification block B onto the metal guide pin
C, pushing it till the notch in the guide pin appears at the front.
• Fix the purification block B by fitting the metal blocking plate D into the groove in the
guide pin C, and lower the cover of the support bracket A.
• Take the sterilizer out of the casing, if any. Place it in an appropriate place and remove
the cover.
• Replace the hose holders of the automatic water inlet 29 joint with the straight fittings
supplied and connect the pipe from support bracket A to the external one.
Speedy Water