• Replace the automatic water inlet pump 46 and the connection pipes with the solenoid
valve EV7 (mounted on its support and with the outlet hose holders and the inlet L-joint)
and the pipes supplied. Use the teflon pipe to connect the internal joint on the back of
the steriliser with the solenoid valve EV7 (inlet 1) and use the silicone pipe to connect the
solenoid valve EV7 (outlet 2) to the clean water tank 10:
• Insert the probe support E in the silicon pipe and fasten the ends with the pipe clamps
provided. Connect to the pins using the appropriate screws and ready to use and
clamped blue cables with eyelets to the filler 14. Clamp probe support E to the supplied
support screwed into the slit J, making sure there is no contact between the pins and
other parts of the steriliser.
• Open the water tap.
• Test the steriliser and the joints.
• Check operations, close the mains water tap and install the sterilizer again.