Viper Low overrun Service Manual
Publication Number: 621360241LOEWSE - 42 - © 2016, Cornelius Inc.
Sanitizing the Barrel
Sanitize the syrup system and barrel by performing the procedure shown in Table 43.
Flushing the System
Flush the wash water from the system by performing the procedure in Table 44.
!
CAUTION:
Flush the system thoroughly, residual sanitizing solution left in the system may create a health hazard.
11.
Perform inspection and replacement of scraper blades. See “Inspecting and
Replacing Scraper Blades” on page 38.
12. Perform a leak test on the barrel. See “Motor Seal Leak Test” on page 46.
Table 43.
Step Action
1.
Use a clean 5-gallon pall filled with a sanitizing solution and water at a tem-
perature of 90º F to 110º F (32º C to 43º C). The sanitizing solution must have
200-ppm available chlorine.
2.
Connect a sanitizing fitting (p/n cc 28688) to the BIB connector. Put the con-
nector in the bucket of sanitizing solution.
3. Make sure the Product/BRIX valve in the BRIX position.
4.
Hold waste container under the BRIX tube to collect flush water from the syrup
line and open the manual syrup flow valve to start filling the syrup line with
sanitizing solution. Continue to hold the syrup flow valve open until sanitizing
solution starts coming out of the BRIX tube.
5. Turn the Product/BRIX valve in the Product position.
6.
Manually override (open) the syrup flow valve to fill the barrel with sanitizing
solution.
7.
Fill the barrel with sanitizing solution by opening the faceplate relief valve until
sanitizing solution comes out of the relief port.
8.
Hold a 16 oz. cup under the dispense valve. Hold the dispense valve fully
open until the cup is full.
9.
Use the brush provided with the unit to clean the relief port and clean the out-
let of the dispense valve with sanitizing solution.
10.
From the Barrel Status menu, press the SPIN button while highlighting the
appropriate barrel. This starts the scraper blade. Allow blade to operate for
minimum of 10, but no more than 15 minutes. Turn the scraper blade off by
pressing the OFF button.
11.
Place a large container under the dispense valve and drain as much sanitizing
solution as possible from the barrel.
12.
When the pressure in the barrel drops, from the Barrel Maintenance menu,
press the PURGE button to re-pressurize the barrel with CO
2. As sanitizing
solution level lowers in the barrel, partially close the valve to avoid spurting.
Table 42.
Step Action