Setting Up the Generator
Setup, Tests, and Adjustment
Force FX Electrosurgical Generator C Service Manual 5-3
6. Turn on the generator by pressing the power switch on (|). Verify the following:
•All visual indicators and displays on the front panel illuminate.
Activation tones sound to verify that the speaker is working properly.
7. If the self-test is successful, a tone sounds. Verify the following:
Indicators above the default mode buttons (Standard bipolar, Pure cut, and
Fulgurate coag) illuminate green.
•Each display shows a power setting of onewatt.
The REM Alarm indicator illuminates red.
If the self-test is not successful, an alarm tone sounds. A number may momentarily
appear in the Cut display and, in most cases, the generator is disabled. Note the
number and refer to Responding to System Alarms on page 6-11.
Important
If the coag mode has been optionally changed to default to Desiccate or Spray, that
corresponding indicator illuminates after the self-test is performed successfully.