Control Board Replacement
7-4 Force FX Electrosurgical Generator C Service Manual
Control Board Replacement
Equipment required
Phillips screwdriver
1. Turn off the generator. Disconnect the power cord from the wall receptacle.
2. Remove the five screws that secure the cover to the chassis. Lift the cover off the
chassis. Set the cover and screws aside for reinstallation.
3. Unlock the connector on the Control board and disconnect the Display board ribbon
cable from the Control board.
4. Carefully slide the Control board straight up through the slots on the heat sinks to
disconnect it from the Power Supply/RF board.
5. Verify that the packaging for the new Control board contains the part number for the
Force FX Electrosurgical Generator C Control board. Do not install any other Control
board.
6. Install the battery in the battery socket (positive side facing up).
7. Position the new Control board over the Power Supply/RF board with the Control
board components facing the rear panel. Fit the edges of the board into the slots on
the heat sinks. Slide the board down, carefully fitting the edge connector into the
matching connector on the Power Supply/RF board.
8. Connect the Display board ribbon cable to the Control board and lock the connector.
9. Position the cover above the chassis and slide it down. Install the five screws that
secure the cover to the chassis.
Warning
Electric Shock Hazard To all o w s t ored en e rgy to diss ipat e after power is disconnected, wait at
least five minutes before replacing parts.
Caution
The generator contains electrostatic-sensitive components. When repairing the generator, work at
a static-control workstation. Wear a grounding strap when handling electrostatic-sensitive
components, except when working on an energized generator. Handle circuit boards by their
nonconductive edges. Use an antistatic container for transport of electrostatic-sensitive
components and circuit boards.
Notice
Calibrate the generator after you install a new Control board. Otherwise, the default calibration
values are used. Refer to Calibrating the Generator on page 5-26, for instructions.