Periodic Safety Check
Setup, Tests, and Adjustment
Force FX Electrosurgical Generator C Service Manual 5-15
Here is a summary of safety checks:
Inspect the generator and accessories
•Inspect the internal components
Test the generator
•Verify REM function
•Confirm outputs
Check leakage current and ground resistance
Recommended Test Equipment
Digital voltmeter—Fluke 77 or 87, or equivalent
•True RMS voltmeterFluke 8920, or equivalent
Oscilloscope—Tektronix 2445, or equivalent
•Leakage current testerUse UL load device or commercially available leakage tester
•Leakage tableper IEC 60601-2-2, Figure 104
•100, 200, 300, 500:, all 250 watt, 1% tolerance, noninductive (Dale NH-250 or
equivalent)
Caution
The generator contains electrostatic-sensitive components. When repairing the generator, work at
a static-control workstation. Wear a grounding strap when handling electrostatic-sensitive
components, except when working on an energized generator. Handle circuit boards by their
nonconductive edges. Use an antistatic container for transport of electrostatic-sensitive
components and circuit boards.