Setting Up the Generator
Setup, Tests, and Adjustment
Force FX Electrosurgical Generator C Service Manual 5-5
Bipolar or macrobipolar connection—handswitching instrument
Setting the Bipolar Output
1. (Optional) To display the previous settings, press the Recall button.
2. To set t he B ipolar mode , press th e Pre cise, Sta nd ard , or Ma cro butt on . The
corresponding indicator illuminates green.
3. To in crease the p ow er for th e selected mode, press the white up arrow ( ' ) button.
To decrea se t he powe r, press the w hi te down arro w ( ) button. The maximum
power setting is 70 W.
Caution
Set power levels to the lowest setting before testing an accessory.
Handswitching instrument