Correcting Battery-Backed RAM Malfunctions
Troubleshooting
Force FX Electrosurgical Generator C Service Manual 6-25
5. Install the Control board.
a. Position the Control board over the Power Supply/RF board with the components
facing the rear panel. Fit the edges of the board into the slots on the heat sinks.
Slide the board down, carefully fitting the edge connector into the matching
connector on the Power Supply/RF board.
b. Connect the Display board ribbon cable to the Control board and lock the
connector.
6. Calibrate the generator. Refer to Calibrating the Generator on page 5-26, for
instructions. If the calibration is successful, install the cover (refer to step 9).
7. If the alarm number reappears, replace the Control board.
a. Turn off the generator.
b. Remove the Control board (refer to step 3).
c. Install the new Control board (refer to step 6).
8. Recalibrate the generator. If the alarm number reappears, record the number and call
the Covidien Service Center.
9. To in stall t he c over, posit ion the c over abov e the chassis and slide it down. Install the
five screws that secure the cover to the chassis.