Kangaroo™ Console | 45
Electrical Safety
The Kangaroo feeding tube with IRIS technology has been built and tested according IEC 60601 Ed. 3.1 2012, ANSI/AAMI ES 60601-1:2005/(R)2012, CAN/CSA C22.2 No. 60601-1-08, IEC 60601-1-6:2013, IEC
60601-1-2 Ed. 3.1:2015 Standards.
The Kangaroo feeding tube with IRIS technology, when used with the console as a system is intended for use in the electromagnetic environment specified below. The user should assure that it is used in such
an environment.
Electromagnetic Conformity Declaration.
Specifications
Section VI: Additional Kangaroo Feeding Tube with IRIS Technology Information
For any questions please contact Covidien Customer Service at 1-800-962-9888, or your local sales representative.
Emissions Tests Compliance Electromagnetic Environment - Guidance
RF emissions (CISPR 11:2015) Group 1 See guidance comments for emissions tests in the tables on the following
pages.
RF emissions (CISPR 11:2015) Class A
Harmonic emissions (EN 60601-3-2:2014) Class A
Voltage Fluctuations / flicker emissions (EN61000-3-3:2013) Complies
Immunity Tests Compliance Electromagnetic Environment - Guidance
Radiated Disturbance Immunity (EN60601-1-2:2015 / IEC 61000-4-3:2010) Complies See guidance comments for immunity tests in the tables on the following
pages.
Conducted Disturbance Immunity (EN60601-1-2:2015 / IEC 61000-4-6:2013) Complies
Power Frequency Magnetic Field Immunity (EN60601-1-2:2015 / IEC 61000-4-8:2009) Complies
Voltage Dips and Sags Immunity (EN60601-1-2:2015 / IEC 61000-4-11:2004) Complies
Electrical Fast Transient / Bursts Immunity (EN60601-1-2:2015 / IEC 61000-4-4:2012) Complies
Electrostatic Discharge Immunity (EN60601-1-2:2015 / IEC 61000-4-2:2008) Complies
Surge Immunity (EN60601-1-2:2015 / IEC 61000-4-5:2014) Complies