Kangaroo™ Console | 46
Section VI: Additional Kangaroo Feeding Tube with IRIS Technology Information
Guidance and manufacturer’s declaration – electromagnetic emission
The Kangaroo feeding tube with IRIS technology, when used with the console as a system is intended for use in the electromagnetic environment specified below. The customer or the user should assure
that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Kangaroo feeding tube with IRIS technology, when used
with the console as a system uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
RF emission
CISPR 11
Class A
The Kangaroo feeding tube with IRIS technology, when used with
the console as a system is suitable for use in all establishments
other than domestic and those directly connected to a low
voltage power supply network which supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies