Kangaroo™ Console | 47
Section VI: Additional Kangaroo Feeding Tube with IRIS Technology Information
Guidance and manufacturer’s declaration – electromagnetic immunity
The Kangaroo feeding tube with IRIS technology, when used with the console as a system is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it
is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnet environment - guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
±2 kV for power supply lines
1kV for signal line
±2 kV for power supply lines
1kV for signal line
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode ±1 kV differential mode Mains power quality should be that of a typical commercial or
hospital environment.
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
0% U
T
(100% dip in U
T
) for 0.5 cycle (50/60 Hz)
0% U
T
(100% dip in U
T
) for 1 cycle (50/60 Hz)
70% U
T
(30% dip in U
T
) for 30 cycles (60 Hz)
70% U
T
(30% dip in U
T
) for 25 cycles (50 Hz)
0% U
T
(70% interruption in U
T
) for 250 cycles (50 Hz)
0% U
T
(70% interruption in U
T
) for 300 cycles (60 Hz)
0% U
T
(100% dip in U
T
) for 0.5 cycle (50/60 Hz)
0% U
T
(100% dip in U
T
) for 1 cycle (50/60 Hz)
70% U
T
(30% dip in U
T
) for 30 cycles (60 Hz)
70% U
T
(30% dip in U
T
) for 25 cycles (50 Hz)
0% U
T
(70% interruption in U
T
) for 250 cycles (50 Hz)
0% U
T
(70% interruption in U
T
) for 300 cycles (60 Hz)
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the Kangaroo feeding tube
with IRIS technology, when used with the console as a system
requires continued operation during power mains interruptions,
it is recommended that it be powered from an uninterruptable
power supply or battery.
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
30A/min 30A/min Mains power quality should be that of a typical commercial or
hospital environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.