EasyManua.ls Logo

Crowcon Gas-Pro - Calibration after Bump Fail

Crowcon Gas-Pro
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
3.2.3 Calibration after bump fail
If any channel fails speedy bump or smart bump then Gas-Pro can be congured (via Portables Pro)
to perform a ‘calibration after bump fail’ immediately after the test failure.
i
Calibration should only be performed with appropriately accurate gas.
As this test immediately follows a speedy or smart bump, if ‘calibration after bump fail’ is congured
the speedy or smart bump test will need to undertaken with calibration quality gas.
i
Ensure the gas applied matches the conguration settings in the Gas-Pro or the test will
fail. This can be done via Portables Pro.
i
Should a ammable sensor be tted, check the label for the original calibration target.
3.2.3.1 Procedure
If calibration is intended, the instrument should have been zeroed manually within 15 minutes
prior to the calibration attempt.
Following a bump test failure, leave the instrument in the Q-test module or with the ow plate
attached and the gas on.
Wait for the calibration result screen indicated by the
or .
The Gas-Pro then returns to normal operation.
During this process the new calibration values are stored to the instrument memory and the
calibration dates are advanced by the congured interval – commonly 1 month as the Gas-Pro has
not been through a formal service/calibration routine (dependant upon region/setting).
If the gas test calibration fails this may be indicative of a more serious sensor issue, including the
need to replace sensors. The instrument should then be serviced.
3.3 New sensor calibration/service
Servicing or the tment of a new sensor can only be undertaken by a suitably trained technician
using the PC software and the appropriate gases.
In addition calibration should be performed as required by local or organisational regulations. In
the absence of suitable evidence, such as a eld assessment by a competent person, Crowcon
recommend regular service and calibration every 6 months.
Contents
Exit
Prologue
Operation
Gas testing
Maintenance
PC software
Accessories
Specication
Troubleshooting
Appendices
Warranty
Set-up

Other manuals for Crowcon Gas-Pro

Related product manuals