Crowcon Gas-Pro - Fault Codes

Crowcon Gas-Pro
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
Fault
codes
Symtom/Error
Message
Icon Cause Action
N/A Bump Test Fail
Lock.
The monitor has not
been gas tested in the
dened period and the
gas test lock function
has been activated.
The monitor needs
calibration.
71 Battery low
Battery Low (about 20
to 30 minutes before
switch off)
Exit the hazardous area
as soon as possible and
recharge battery.
9.2.2 Fault codes
The following fault codes are not xable by the user and the instrument should be sent to an
authorised service agent:-
Fault code 4,5,6,7,8,11,12,13,14,15,16,17,19,20,25,26,27,28,29,30,50,51,52,53,54.