8. Specication
Detector type Multi-gas monitor
Gases* O
2
, H
2
S, CO, CO
2
, FLAM Pellistor (CH
4
, C
5
H
12
, C
3
H
8
, C
4
H1
0
, C
2
H
4
, C
2
H
2
, H
2
), SO
2
, CL
2
,
CLO
2
, NO, NO
2
, NH
3
, O
3
, PID, FLAM IR (CH
4
, C
5
H
12
, C
3
H
8
)
Size (d x l x w) 43 x 130 x 84 mm (1.7 x 5.1 x 3.3 inches)
Weight 5 gas (pump) 362g (12.7oz)
5 gas (un pumped) 333g (11.7oz)
4 gas (pump) 340g (11.9oz)
4 gas (un pumped) 309g (10.8oz)
Alarms Audible>95dB
Visual – all angle dual red/blue LEDs
Vibrating alert
+ve Safety™
Display Top mount for ease of view
viewable size 25 x 50 mm
Data logging 125hrs @10 second intervals (45,000 logs)
Event logging Alarm, over range, calibration, bump, on/off, TWA, 1000 events
Battery Rechargeable li-ion
Sampling Internal pump as option
Hand aspirator for non pumped
Operating temperature -20°C to +55°C
Storage -25°C to +65°C (-13°F to +149°F)
Humidity 10 to 95 % RH
Ingress protection Independently tested to IP65 and IP67
Approvals IECEx: Ex d ia IIC T4 Gb Tamb -20°C to +55°C
ATEX:
II 2 G Ex d ia IIC T4 Gb Tamb -20°C to +55°C
UL: Gas detector use in hazardous locations Class 1 Division 1
Groups A, B, C and D only as to intrinsic safety
Marine Equipment Directive:
Lloyds Certicate Number : MED 1500009/M2
Compliance CE, FCC and ICES-003
Complies with EMC Directive 2004/108/EC
Interface Data connection for use with calibration stations & direct to PC
Charging
Options
Direct connect to multi-region power supply
Car charging adapter
USB interface cable
USB power & communications cable
5 Way multi-region power supply
* Gases not available in Gas-Pro with the internal pump option are CL
2
, CLO
2
& O
3
.
Sensors may be degraded at the higher temperatures subject to individual sensor specications
Dependant upon sensor conguration
Gas-Pro’s tted with a PID sensor are rated to IP65 only