EasyManuals Logo

CT Lab Vecto III User Manual

CT Lab Vecto III
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #34 background imageLoading...
Page #34 background image
Vecto III User Manual
2.4.4 Voltage Inputs
The Vecto III has 4 resistive differential voltage inputs made up of 6 terminals.
The first 4 terminals are used for 3 voltage inputs. V
1
, V
2
and V
3
are Neutral (V
com
)
referenced (V
1
-N, V
2
-N, V
3
-N).
The last 2 terminals are used for the 4
th
(V
4+
, V
4-
) voltage input. It is a full differential
measurement normally used for Neutral to Earth measurement.
The voltage inputs make use of a 6-terminal lockable connector plug. Each terminal
can accommodate conductors with sizes up to 2.5 mm
2
. Figure 9 shows the voltage
connections for a 3-phase 4-wire star system. Power is calculated per phase and
arithmetically added to obtain the 3-phase powers.
Figure 10 shows the voltage connections for a 3-phase 3-wire delta connection. Three-
phase power is calculated by the 2-watt meter method.
If voltages are connected in star (4-wire) on the terminals the
user can select star (4-wire) or delta (3-wire) topology in the
software

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CT Lab Vecto III and is the answer not in the manual?

CT Lab Vecto III Specifications

General IconGeneral
BrandCT Lab
ModelVecto III
CategoryMeasuring Instruments
LanguageEnglish