Vecto III User Manual
2.3 Device Overview
2.3.1 Key Features
The Vecto III records both voltages and currents of a single 3-phase feeder with IEC
61000-4-30 Class A performance.
Voltage inputs:
Class A performance over all of the input range inclusive of 63.5 V (if L-N
measurements are done in a 110 V line-line VT circuit)
6 terminals are used for 4 voltage input channels with maximum input
ratings of 600 V.
o The first 4 terminals connect 3 different phase voltages such as in a
3-phase system.
o Input channel 1-3 is referenced to a neutral (4
th
terminal)
connection. This 4
th
terminal acts as a reference channel and can be
any phase voltage if a 3-phase line-line voltage circuit is measured.
o If it is a 3-phase line-neutral voltage circuit, then the 4
th
terminal
will be connected to the neutral voltage.
o The 5
th
and 6
th
terminal is the 4
th
channel and a fully differential (2
terminal) input channel.
Current inputs:
4 X 1/5 A CT inputs
4 X 1 V
rms
differential voltage transducer inputs with a 5 V power supply
(to be used to power Rogowski coils).
Digitisation of input signals
500 kHz sampling of primary input signals.
50 kHz is digitally retained and subjected to the IEC 61000-4-30 Class A
signal processing requirements.
Power supply
PoE and/or a 90-300 V AC/DC power factor corrected power supply.
1/2 hour power failure ride-through via an internal LiFePO
4
battery.
Data storage