EasyManua.ls Logo

Datacolor Elrepho 3000 - Page 56

Datacolor Elrepho 3000
79 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Elrepho 3000
To prevent the system from “crashing”, no result is computed and
this message is displayed.
SQUARE ROOT OF NEGATIVE #
This message indicates the measurement data passed to the task’s
formula resulted in values which would result in the performing of a
square root operation on a negative number. To prevent the system
from “crashing”, no result is computed and this message is
displayed.
HELMHOLTZ DOMINANT WAVELENGTH RANGE ERR
This message indicates the measurement data passed to the
Helmholtz formula resulted in a dominant wavelength which is
outside of the valid range of 570nm - 585nm.
TASK COMPUTE ERROR
This message indicates the measurement data passed to the task’s
formula resulted in some type of miscellaneous error. To prevent
the system from “crashing”, no result is computed and this message
is displayed.
UNKNOWN SYSTEM ERROR
This message will be displayed if some type of miscellaneous error
occurs during the power up procedure. This may indicate a possible
instrument malfunction. Please contact your Datacolor
International Service representative for further instructions.
WHITE TILE DATA NOT LOADED
This message will be displayed if the instrument’s white tile data is
not found during the power up procedure. This condition could
occur if the white tile data was not loaded with the application
program, or deleted. Please contact your Datacolor International
Service representative for further instructions.
TASK DATA NOT LOADED
This message will be displayed if the instrument’s task data is not
found during the power up procedure. This condition could occur if
the task data was not loaded with the application program, or
56 Datacolor International

Table of Contents

Related product manuals