Accutorr Plus Service Manual 0070-10-0691 2 - 1
2.0
Specifications
2.1 Performance Specifications ........................................................... 2-1
2.2 Safety Characteristics ..................................................................... 2-6
2.3 Physical Characteristics ................................................................. 2-7
2.4 Environmental Characteristics ....................................................... 2-7
2.5 Electrical Ratings ............................................................................. 2-8
2.6 Agency Compliance ........................................................................ 2-8
2.7 Electromagnetic Compatibility ....................................................... 2-9
2.1 Performance Specifications
Systolic and Diastolic Pressure Readout
Number of Digits: 3
Accuracy*: Mean error <
±5 mmHg
Standard deviation <
±8 mmHg
Range: Adult Mode: 55 to 260 mmHg
Pediatric Mode: 55 to 160 mmHg
Neonatal Mode: 45 to 120 mmHg
* Tested per ANSI/AAMI SP10-1992 methods.
2.1.1 NIBP Measurement Cycle Time
Less than 30 seconds average at 80 BPM with 180 mmHg pump up pressure, without
retries, motion artifact, or arrhythmia with standard adult cuff on a healthy individual.
Cycle time is affected by arm size and wrapping technique.