EasyManua.ls Logo

Datascope Accutorr Plus - Overpressure Transducer Verification <12 C, 12 A, 12 B; Overpressure Transducer Calibration

Datascope Accutorr Plus
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Accutorr Plus Service Manual 0070-10-0691 5 - 15
Calibration Service Diagnostics
5.5.16 Overpressure Transducer Calibration
NOTE: This procedure should only be performed if a
failure occurred in steps 1 through 3 of 5.5.14
“Overpressure Transducer Verification <12c, 12a,
12b>” on page 5-13.
This is the procedure for calibrating the overpressure transducer in the event that the
display does not read 158 mmHg during the test for Low pressure mode in 5.5.14
“Overpressure Transducer Verification <12c, 12a, 12b>” on page 5-13. The transducer is
calibrated using R504 on the 0670-00-0814 PCB (see Figure 5-5).
1. Adjust R504 on the 0670-00-0814 PCB until the MAP readout is 158 mmHg.
2. Repeat 5.5.14 “Overpressure Transducer Verification <12c, 12a, 12b>” on page 5-13.
FIGURE 5-5 Overpressure Adjust (R504)

Table of Contents

Related product manuals