Accutorr Plus Service Manual 0070-10-0691 3 - 5
Repair Information Troubleshooting (Problem Isolation)
3.5 Troubleshooting (Problem Isolation)
This section has been divided into two major areas: clinical and technical. Clinical
problems can generally be resolved by the user. Technical problems may require
resolution by a qualified technical service person. The error code table below indicates
Technical problems with an asterisk (*). Error codes are displayed on the front panel
LEDs of the Accutorr Plus.
3.5.0.1 Error Codes
TYPE CODE DESCRIPTION REASON
NIBP 8810 Retry - Unable to Measure Motion artifact, cycle time-out,
weak pulsations or no pulsations.
A triple beep tone is generated.
8811 Retry - Pump Higher Insufficient cuff pressure. A triple
beep tone is generated.
8812 Stop - Cuff Overpressure Excessive cuff pressure detected
by the software. A triple beep tone
is generated.
8813 Stop - Unable to Measure 4 successive measurement
attempts failed. A triple beep tone
is generated.
TEMP
(PTM)
8830 Check Probe Tissue contact may have been
lost.
8831 Replace Probe Defective probe or connection.
SpO
2
8850 No Sensor No sensor connected.
8851 Sensor Off Sensor not on patient.
(Masimo
®
SpO
2
only)
8852 Interference Interference on signal.
(Masimo
®
SpO
2
only)
8853 Pulse Search Unit cannot find signal. (Nellcor
®
SpO
2
module will report ‘‘Pulse
Search" -8853- when the sensor
is not on the patient.)
8854 Weak Pulse Weak pulse detected.
(Masimo
®
SpO
2
only)
8856 Check Sensor Sensor problem.
(Masimo
®
SpO
2
only)
8857 PR<21 Pulse rate is less than 21 bpm.
(Nellcor
®
SpO
2
only)
8857 PR<26 Pulse rate is less than 26 bpm.
(Masimo
®
SpO
2
only)
8858 PR>249 Pulse rate is greater than 249
bpm. (Nellcor
®
SpO
2
only)
8858 PR>239 Pulse rate is greater than 239
bpm. (Masimo
®
SpO
2
only)