Accutorr Plus Service Manual 0070-10-0691 5 - 5
Calibration Service Diagnostics
5.5 Service Diagnostics
The Service Diagnostics tests must be run in the sequence indicated. The numbers and
letters in parentheses after the title of each test, represent the room number and bed letter
used to access the test.
5.5.1 Introduction (Hidden Key)
To enter the Service Diagnostics mode, while powering on the unit, press and hold the
hidden key until two beeps are heard. The hidden key is the Adult symbol above the
Patient SETUP key.
Verify that after the initial power up sequence a “0” is displayed in the Room LED, “a” is
displayed in the Bed LED, and “ddd” is displayed in the Systolic LEDs.