EasyManua.ls Logo

Datascope Accutorr Plus - Physical Characteristics; Environmental Characteristics

Datascope Accutorr Plus
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Accutorr Plus Service Manual 0070-10-0691 2 - 7
Specifications Physical Characteristics
2.3 Physical Characteristics
Size (maximum):
Main Unit: 19 cm(W) x 26.93 cm(H) x 20.83 cm (D)
7.5" (W) x 10.6" (H) x 8.2" (D)
Recorder Module: 5.33 cm(W) x 23 cm(H) x 11 cm (D)
2.1" (W) x 9" (H) x 4.25" (D)
Predictive Module: 5.7 cm(W) x 15.9 cm(H) x 11.8 cm (D)
2.25" (W) x 6.25" (H) x 4.63" (D)
Weight: <4.95 kg (11 pounds), depending on configuration
2.4 Environmental Characteristics
Operating
Temperature: 10°C to 40°C, 50°F to 104°F (Accutorr Plus
& Recorder)
10°C to 32°C, 50°F to 90°F (Predictive
Thermometer)
Humidity: 15 to 90% max, non-condensing.
Altitude: 1013 hPa to 697 hPa (0 to 10,000 ft.)
Shock and Vibration: Meets IEC 60068-2-27 for shock with peak
acceleration of 15g, 11 mSec duration, half sine.
Meets EN 60068-2-64 for random vibration
with frequencies of 10 to 2000 Hz, resolution of
10Hz, 10 minutes per axis, acceleration spectral
density of:
10 Hz to 100 Hz : 1.0 (m/s2)2/Hz
100 Hz to 200 Hz : -3 dB per octave
200 Hz to 2000 Hz : 0.5 (m/s2)2/Hz
Shipping: Meets ISTA Test Procedure 1A (less than 100 lbs)
Storage
Temperature: -15°C to +60°C
+5°F to 140°F
Humidity: 10 to 95%, non-condensing

Table of Contents