314Appendix F: Technical Information
Table 7-B. Sensor Survival Rate by Wear Day (Pediatrics; n=210)
Wear Day Number of Sensors Survival Rate (%)
1 206 99.0%
2 204 99.0%
3 196 97.1%
4 193 95.6%
5 184 91.1%
6 175 88.6%
7 164 85.5%
8 157 83.4%
9 146 79.2%
10 142* 76.8%
* Includes sensors that survived more than 9.5 days (228 hours) of wear.