322Appendix F: Technical Information
Transmitter Performance Characteristics
Parameter Performance Characteristic
TX/RX Frequencies 2.402 – 2.480 GHz
Bandwidth 1.07 MHz
Maximum Output Power 1.0 mW EIRP
Modulation Gaussian Frequency-Shift Keying
Data Rate 1 Mbps
Data Communication Range 20 feet
The G6 is an M-PED (Medical-Portable Electronic Device), which meets the FAA RTCA /
DO-160 edition G Section 20 Category T and Section 21, Category M. It can be used on
aircraft according to the directions provided by the operator of the aircraft.
This device can withstand exposure to common electrostatic discharge (ESD) and
electromagnetic interference (EMI).
Quality of Service Summary
Quality of Service for the G6 System wireless communication using Bluetooth Low
Energy is assured within the effective range of 20 feet, unobstructed, between the G6
transmitter and paired display device at regular 5-minute intervals. If connection is lost
between the transmitter and display device, upon re-connection any missed packets
(up to 3 hours) will be transmitted from the transmitter to the display device. The G6
CGM System is designed to only accept radio frequency (RF) communications from
recognized and paired display devices.