324Appendix F: Technical Information
Transmitter Electromagnetic Immunity Specifications
Immunity Test
IEC 60601 Test
Level
Transmitter
Compliance
Level
Electromagnetic
Environment Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
± 8 kV Contact
± 15 kV Air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Power Frequency
(50/60 Hz)
Magnetic Field
IEC 61000-4-8
30 A/m 30 A/m
Magnetic fields from
common appliances are
not expected to affect the
system.
Receiver Product Specifications
Part Number MT24078
Reading Frequency Every 5 minutes
Dimensions
Length: >4.2 inches
Width: >2.5 inches
Thickness: >0.55 inches
TX/RX Frequencies 2.402 – 2.480 GHz
Bandwidth 1.39 MHz
Maximum Output Power 2.4 mW EIRP
Modulation Gaussian Frequency-Shift Keying