DIESTRO 103APV4R ELECTROLYTE ANALYZER
39 JS Medicina Electrónica
Route: "HOME" screen → Calibration → Options → Corrections
From the drop-down list, select the electrode to which you want to apply the correction. Then
press "a" to modify the quadratic term, "b" for the linear term, and "c" for the delta.
Corection equation:
y = ax
2
+ bx + c
The Correction equation is typical of the equipment and is maintained when
changing the Pack or electrodes. Verify the need to maintain correction values
with the change of the pack or electrodes.