DIESTRO 103APV4R ELECTROLYTE ANALYZER
66 JS Medicina Electrónica
EXAMPLE OF STRINGS SENT OVER THE NETWORK:
- Serum measurements in analyzer with Na-K-Cl-Ca-Li configuration, Sample number 7,
Patient Juan Perez1234, Na Measurement with errors ? Cl, K Measurement with errors ? Q, Cl
Measurements OK, Ca below normal value and Li above normal value.
All ions are in units of mmol.
# 7;&Juan Perez1234 ;2020/11/09 11:09:39; Na=145.3mmol/E:u ; K = 4.16mmol/E:Q ;
Cl=105.7mmol ; Ca= 0.85mmolv ; Li= 0.51mmol^;;;
- Urine measurements in the analyzer with Na-K-Cl-Ca-Li configuration, Urine
measurements without errors, without patient ID. All units of the ions are in mmol.
# 8;&;2012/10/13 16:39:57; NaO=251.0 mmol/l;KO=21.20 mmol/l;ClO=251.0 mmol/l;;;
Note: In urine measurements, a letter "O" is added after the measured ion. Only Sodium,
Potassium and Chlorine are reported.
SENDING RESULTS FROM MEMORY TO THE NETWORK
To send a result from memory to the Network, look for the desired result in memory (See
section 12 - Storage of Results) and press the button for sending data: