DIESTRO 103APV4R ELECTROLYTE ANALYZER
65 JS Medicina Electrónica
▪ The analyzer sends the measurement data in a data string with the following format:
# Sample Number;& Patient; Date Time; Na measurement result; Measurement result of
K; Cl measurement result;Ca measurement result; Li;CR measurement result
Increase from 1 to 65535. Not editable by the user.
Allows up to 20 alphanumeric characters entered by
keyboard, barcode or keyboard built into the analyzer
Between date and time, there are 2 separation spaces.
Serum Na
measurement result
Z: Measurement Error Flag (/? SCLQuU) v^: Result
outside the established normal range
Calcium units can be mmol, mgr % or meq/l
W W measurement
result in serum
Serum Cl
measurement result
Serum Ca
measurement result
Result of Li
measurement in
serum
Result of the
measurement of Na
in urine
NaO=XXX.X mmol v^ Error:Z
Result of the
measurement of K in
urine
KO= XX.XX mmol v^ Error:Z
Result of the
measurement of Cl in
urine
ClO=XXX.X mmol v^ Error:Z
Carriage Return (String Completion)
Only sample results are sent through the port. The results of Quality Control measurements or
Calibrations will not be sent.