Weight readings don't seem
to be correct.
Check the scale's accuracy with a test weight.
Recalibrate if necessary.
Scale drifts off of zero. Check for air currents and/or vibration around the
scale. If that is the cause it may be necessary to
set the AZT aperture to a wider setting to
compensate (see Section 4).
Scale reading is bouncing or
"flighty".
Check for air currents and/or vibration around the
scale. If that is the cause it may be necessary to
set the Digital Averaging to a higher setting to
stabilize the reading (see Section 4).
If you are still experiencing a problem with your7000XL, or if the problem you are
having is not covered in the above list, please contact your Doran Scales authorized
dealer.
Resetting the scale parameters:
If at some point the Model 7000XL or Model 7000, user wishes to return
the setup parameters to factory default, follow these steps.
Remove power.
Place scale in the calibration mode and hold the ZERO push button
while power is restored.
The indicator will display "InIt" until the ZERO button is released. After
ZERO is released, the scale will perform its normal power up routine
and then enter the Calibration mode. At this time, all the parameters
will have been reset to their factory default settings. See Section 4 for
details on setting up the individual scale parameters.
Return the switch to the normal weighing position. The scale will save
the revised parameters and will enter the normal weighing mode.
Resetting the scale:
In the event that a power problem has disabled the scale, remove power,
wait 15 seconds and restore power. The scale should restart and function
properly.