Alarms and Messages
VF4 Infinity Gamma Series Page 5-3
Serious Alarms
A serious alarm has the second highest priority. It alerts you to
significant changes in the patient’s condition other than life-
threatening events as defined in the preceding section. For exam-
ple, if a monitored parameter falls below the selected alarm lim-
its, the monitor triggers a serious alarm. An unrecognized
physiological condition, as when a detected parameter is outside
the monitors measuring range, triggers a serious alarm also.
NOTE: Apnea events trigger a serious alarm in all monitoring
modes.
A serious alarm has the following characteristics:
! The parameter box blinks and appears yellow.
! The monitor produces two short tones continuously.
! The monitor displays the alarm cause in both the parameter
box and the message area.
! The monitor stops the alarm automatically when the condi-
tion ceases to exist.
! The monitor stops the alarm when a new alarm of equal or
higher grade is triggered.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals