Page 11-14 Infinity Gamma Series VF4
Rsp Safety Considerations
WARNINGS:
! Do not rely upon impedance respiration monitoring
as the sole method for detecting the cessation of
breathing. Patients at risk for respiratory distress
should be observed closely and heart rate alarms
should be enabled and set appropriately. The use of
additional methods for respiratory monitoring, such
as etCO
2
and SpO
2
are recommended whenever pos-
sible.
! Large amplitude pacer spikes (100 mV or greater)
may interfere with the monitor’s ability to monitor
respiration.
! Impedance respiration monitoring and pacemaker
spike detection are inoperative when using either the
shielded blue leads or the ESU block.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals