Respiration Monitoring
VF4 Infinity Gamma Series Page 11-17
OCRG Recordings
The monitor can print OCRG alarm and manual recordings, if
OCRG waveforms are displayed. OCRG recordings show the
second and the third waveform channels (not the first channel).
An apnea event triggers an OCRG alarm recording, which
includes 108 seconds of pre-apnea OCRG data and 36 seconds of
post-apnea OCRG data.
If no recorder is available, the monitor stores the OCRG apnea
alarm recording (see the chapter Recordings for a description of
stored recordings).
NOTE: OCRG recordings cannot be printed on a laser printer.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals