Page 10-8 Infinity Gamma Series VF4
ST Alarms
Set ST alarm limits on the Alarm Limits table (see the chapter
Alarms and Messages). The default ST segment deviation alarm
limits are +1.0 mm (or +0.10 mV) for the upper limit and -1.0
mm (or -0.10 mV) for the lower limit. However, adjustments to
the upper and lower alarm limit settings must be made based on
careful clinical evaluation. You can set the upper and lower lim-
its within the following range:
! -15.0 mm to +15.0 mm, in 0.1 mm increments (-1.5 mV to
+1.5 mV, in 0.01 mV increments, when using millivolts.)
NOTE: Always check the upper and lower ST alarm limits before
starting ST monitoring.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals