Page B-12 Infinity Gamma Series VF4
Cleaning/Replacing the Fan Filter
The fan filter should be cleaned once per month and replaced at
least once a year.
STEPS: Cleaning/Replacing the Fan Filter
1. Grasp the fan filter and remove it from its holding slots on the
back of the module.
2. Vacuum up any accumulation of dust at the fan port and
inside the filter (if reusing the filter).
3. Reinsert the cleaned fan filter or a new filter.
Fan Filter