Page 8-4 Infinity Gamma Series VF4
3. Remove the outer epidermal layer as required to reduce skin
impedance. Mildly abrade only the electrode contact site with
ultrafine sandpaper (220-400 grit). Apply the electrodes one
at a time and make sure the electrode gel is in contact with
the abraded skin area.
4. For severely diaphoretic patients, use a benzoin prep for
tighter adherence of the electrodes.
5. Inspect the electrode gel to make sure it is moist. Apply the
pad with a circular motion on the adhesive area first, then
press gently on the gel area to prevent the gel from being
squeezed out.
6. Change electrodes every 24-48 hours to ensure a good quality
signal. However, if the electrocardiographic pattern becomes
less distinct, if the patient is diaphoretic, or if skin irritation
develops, the electrodes must be changed and reapplied
sooner.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals