EasyManuals Logo

Dräger Infinity Series User Manual

Dräger Infinity Series
384 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #281 background imageLoading...
Page #281 background image
Non-Invasive Blood Pressure
VF4 Infinity Gamma Series Page 15-9
STEPS: Selecting the Inflation Mode
1. Click on the NBP parameter box.
2. Click on
Inflation Mode.
3. Select the desired inflation mode and click the knob.
WARNIN G: In neonatal monitoring mode, a blood pres-
sure higher than the inflation range may activate the
NBP Cannot Measure”, “NBP No Pulsation”, or “NBP
Mean Only” alarms. If this happens, manually check the
patient’s pressure and, if appropriate, select the Neo 140
inflation mode or switch to a Pediatric inflation mode.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)